TB 9-6625-2285-40
(3) Connect resistance standard No. 1, 1 MΩ, HI terminal to TI TANK UNITS HIZ
terminal and LO terminal to TI TANK UNITS LOZ terminal, using special test cable.
(4) TI display will indicate between .9000 and 1.100.
(5) Set SELECT switch to COMP-HIZ.
(6) Move resistance standard No. 1 connection from TI TANK UNITS LOZ to TI
TANK UNITS COMP (replace special cable).
(7) TI display will indicate between .9000 and 1.100.
(8) Set SELECT switch to LOZ-COMP.
(9) Move resistance standard No. 1 connection from TI TANK UNITS HIZ to TI
TANK UNITS LOZ.
(10) TI display will indicate between .9000 and 1.100.
(11) Set SELECT switch to LOZ-SHLD.
(12) Move resistance standard No. 1 connection from TI TANK UNITS COMP to
SHLD of TI TANK UNITS LOZ.
(13) TI display will indicate between .9000 and 1.100.
NOTE
GND, when indicated, means CHASSIS connection.
(14) Set SELECT switch to each remaining position and move resistance standard
No. 1 connections as indicated by SELECT switch position.
(15) TI display will indicate between .9000 and 1.100 for each SELECT
switch position.
(16) Set SELECT switch to HIZ-SHLD position.
(17) Connect resistance standards as indicated in table 3 to TI TANK UNITS HIZ
and TI TANK UNITS HIZ outer shell (SHLD). TI indications will be within those listed
Table 3. Resistance Measurement Accuracy
Resistance
Test instrument
standards settings
Min
Max
Adjustments
(MΩ)
10
9.00
11.00
b(1) through (5) and (10) (R)
100
90.00
110.00
b(1) through (3) and (6), (7), and (10) (R)
1000
900.00
1100.00
b(1) through (3) and (8) through (10) (R)
10,0001
9000.00
11000.00
b(1) through (3) and (10) through (12) (R)
1Substitute
resistance standard No. 1 with resistance standard No. 2.
8