TB 9-6625-2285-40
b. Adjustments
(1) Remove resistance standard connections.
(2) Remove six screws from front panel and lift front panel from case.
(3) Set SELECT switch to LOZ-HIZ.
(4) Connect 10 MΩ resistance standard No. 1, to TI TANK UNITS LOZ and HIZ
using special test cables.
(5) Adjust MLRZ potentiometer (fig. 2) for display indication of 10.00 (R).
(6) Connect 100 MΩ resistance standard No. 1, to TI TANK UNITS LOZ and HIZ
using special test cables.
(7) Adjust MMRZ potentiometer (fig. 2) for display indication of 100.0 (R).
(8) Connect 1000 MΩ resistance standard No. 1, to TI TANKS UNITS LOZ and
HIZ using special test cables.
(9) Adjust MHRZ potentiometer (fig. 2) for display indication of 1000 (R).
(10) Connect 10 GΩ resistance standard No. 2, to TI TANK UNITS LOZ and HIZ
using special test cables.
(11) Adjust MERZ potentiometer (fig. 2) for display indication of 10000 (R).
(12) Replace front panel in case and reinstall six screws.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
9/(10 Blank)