Frequency Response
Power Supply

Calibration Procedure For Electronic Voltmeter An/Urm-145B And An/Urm-145C (Me-247C/U) Page Navigation
  2    3    4    5    6  7  8    9  
img
. . . . . .
TB 9-6625-2289-24
Figure 2. Frequency response - equipment setup.
(7) Adjust signal generator frequency controls to value listed in signal generator
frequency settings listed in table 5.
(8) Configure measuring receiver to measure RF power in mV.
(9) Measuring receiver indications will be within the limits listed in table 5.
(10) Adjust signal generator for remaining frequency settings listed in table 5.
Measuring receiver indication for remaining frequency settings will be within limits
specified in table 5.
Table 5. Frequency Response
Signal
generator
Measuring receiver
frequency
voltage indications
settings
(mV)
(MHz)
Min
Max
3
90
110
5
90
110
7
90
110
10
90
110
20
90
110
50
90
110
90
90
110
120
95
105
200
95
105
300
95
105
400
95
105
8