Figure 2. Frequency response - equipment setup.
Figure 3. Regulated power supply 3A1 - location of parts.

Calibration Procedure For Electronic Voltmeter An/Urm-145B And An/Urm-145C (Me-247C/U) Page Navigation
  3    4    5    6    7  8  9  
img
. . . . . .
TB 9-6625-2289-24
b. Adjustments. No adjustments can be made.
10. Power Supply
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
a. Performance Check. Connect digital multimeter positive lead to 3A1TP1 (fig. 3)
and connect negative lead to 3A1TP2 (fig. 3).  If digital multimeter does not indicate
between +14.49 and +14.51 V dc, perform b (1) below.
b. Adjustments. Adjust 3A1R13 (fig. 3) for a digital multimeter indication between
+14.49 and +14.51 V dc (ideal 14.50 V dc) (R).
11. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
9