TB 9-6625-2293-24
NOTE
When POWER pushbutton is pressed to ON, TI automatically
performs a self-test sequence. Upon successful completion of
self-test, TI will be in normal operating mode.
b. Press POWER pushbutton to ON and allow at least 20 minutes for equipment warm-up.
NOTE
If TI crt displays DIAGNOSTIC, PUSH A/B TRIG TO EXIT
at power on, one of the power-up tests has failed. If the error
message on the bottom line of the display is TEST 04 FAIL xx
where xx is 01, 02, 10, or 11, stored calibration data is in error.
If this error message is displayed, press A/B TRIG pushbutton
to exit diagnostic mode and perform SECTION IV. If any
other error message occurs, the failure is probably not related
to calibration. In this case, the instrument should be repaired
before attempting calibration.
c. Press corresponding pushbutton for indications as listed in (1) through (9) below:
(1) VERTICAL MODE CH 1 and CH 2 on.
(2) VERTICAL MODE CH 3 and CH 4 off.
(3) VERTICAL MODE ADD and INVERT off.
(4) VERTICAL MODE CHOP/ALT to ALT.
(5) 20 MHz BW LIMIT to off.
(6) TRACK/INDEP to INDEP.
(7) TRIGGER SLOPE to + (plus).
(8) ΔV and Δt off (cursors off).
(9) X10 MAG to off (out).
d. Set corresponding switches for indications as listed in (1) through (4) below:
(1) TRIGGER MODE up to AUTO LVL.
(2) TRIGGER SOURCE up to VERT.
(3) TRIGGER COUPLING up to DC.
(4) CH 1 and CH 2 input coupling up to 1 M Ω AC.
e. Position controls as listed in (1) through (10) below:
(1) VERTICAL POSITION to midrange.
(2) CH 1 and CH 2 VOLTS/DIV to 100 mV and VAR control cw to detent.
(3) A SWP and B SWP SEC/DIV to 1 ms (knobs locked).
(4) SEC/DIV VAR cw to detent.
(5) Horizontal POSITION to midrange.
(6) TRIGGER LEVEL to midrange.
5