TB 9-6625-2293-24
Table 2. Minimum Specifications of Equipment Required
Common name
Minimum use
Manufacturer and model
specifications
(part number)
Range: 0 to 20 V
Fluke, Model 8840A/AF05 (AN/GSM-64D)
Accuracy: 0.025%
Fluke, Model 5820A-5C-GHZ (5820A-5C-GHZ)
Volts out:
CALIBRATOR
Range: 10 mV to 20 V
Accuracy: 0.5%
Time markers:
Range: 0.5 ns to 0.5 s
Accuracy: 0.175%
Pulses:
Risetime: 0.225 ns
Range: 1 Hz to 500 MHz
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. When indications specified in paragraphs 7 through 13 are not within tolerance,
perform SECTION IV, MENU CALIBRATION PROCESS. After adjustments are made,
within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
the calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Connect TI to a 115 V ac source.