CAL 01 Horizontal - Continued
Table 15. Vertical Attenuator Check

Calibration Procedure For Oscilloscope Model 2465 Page Navigation
  21    22    23    24    25  26  27    28    29    30    31  
img
. . . . .
TB 9-6625-2293-24
Table 13. Timing (CAL 01 Steps 7 Through 16)
Test
Test instrument
instrument
superimposed
Display
ΔREF
Δt
displayed
Oscilloscope calibrator
step
B sweep
output
number
time marker
marker
division
settings
μS/D
7
2
10
1
0.2
μS/D
8
2
10
2
0.2
μS/D
91
4
28
2
1.2
μS/D
10
2
10
10
0.2
μS/D
11
2
10
50
0.2
μS/D
121
4
28
50
1.2
μS/D
13
2
10
0.5
0.2
μS/D
141
4
28
0.5
1.2
μS/D
22
15
10
0.1
0.2
22
16
10
20
nS/D
0.1
Δ control is adjusted at step 9, 12, or 14 the previous step will be repeated.
1If
2Intensify
starts on indicated time marker and may cover more than 1 time marker.
Table 14. Timing (CAL 01 Steps 17 Through 29)
Oscilloscope
Test instrument
calibrator
Displayed
Time markers
output settings
Step
over 8 divisions
μS/D
17
8
1
μS/D
18
24
1
μS/D
19
8
2
μS/D
20
24
2
μS/D
21
8
10
μS/D
22
8
50
μS/D
23
24
50
μS/D
24
8
0.5
μS/D
25
24
0.5
μS/D
26
8
0.1
27
8
20
nS/D
28
2
2
nS/D
29
8
1
mS/D
NOTE
If Δ control is adjusted at step 18, 20, 23 or 25 the previous
step will be repeated. At step 28, adjust TRANS RESP R802
(fig 1) for precisely two cycles between the 2d and 10th graticule
lines.
(35) After step 29 has been completed display will indicate DIAGNOSTIC PUSH
A/B TRIG TO EXIT.
20. CAL 02 -Vertical
a. Performance Check
(1) Set TRIGGER MODE switch up to step to CAL 02 (bottom left of display).
27