TB 9-6625-2294-35
(26) If cursors are not aligned with 2d and 10th vertical graticule lines, adjust XI
GAIN R860 (fig. 1), and HORZ CTR (R801) (fig. 1) to align the two cursors with 2nd and
10th vertical graticule line (R).
(27) If TI does not display 1 time marker per division l minor division, adjust X10
GAIN R850 (fig. 1) for 1 marker per division (R).
(28) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate: ADJ ∆, (step) 6, 10 ms, 100 s.
(29) Set oscilloscope calibrator MARKER output to 10 S/D and adjust ∆REF OR
DLY POS control to intensify 2d time marker, adjust ∆ control to intensify 10th time
marker, and superimpose delayed sweep time markers within .2 division.
(30) Press and release upper TRIGGER COUPLING pushbutton, and set
oscilloscope calibrator output for 1 mS/D.
(31) For each step in table 20 do the following:
(a) Adjust ∆REF OR DLY POS and ∆ controls, as necessary, to intensify
indicated time markers on the A sweep and superimpose displayed B sweep time markers
within listed limits.
(b) Press and release upper TRIGGER COUPLING pushbutton.
Test instrument
Test instrument
Oscilloscope
superimposed
calibrator
MARKER
Displayed
displayed B sweep
∆ REF
∆
output settings
step number
division
time marker
time marker
7
2
10
0.2
S/D
1
8
2
10
0.2
S/D
2
91
4
28
1.2
S/D
2
10
2
10
0.2
S/D
10
11
2
10
0.2
S/D
50
121
4
28
1.2
S/D
50
13
2
10
0.2
S/D
.5
141
4
29
1.2
S/D
.5
15
32
19
50
nS/D
0.2
16
22
10
20
nS/D
0.1
1If
∆ control is adjusted at step 9, 12, or 14 the previous step will be repeated.
2Intensifies
starts on indicated time marker and may cover more than 1 time marker.
(32) After completion of step 16 of table 20, TI display will indicate: ADJ ∆, (step)
17, 1 s.
(33) Adjust TRACE SEP control fully cw.
(34) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to TI CH 1 and
oscilloscope Vertical 1 input.