TB 9-6625-2295-35
(27) Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to adjust
amplitude for 5 divisions of deflection on TI. If oscilloscope calibrator err display does not
indicate within limits specified in first row of table 6, perform b below.
(28) Repeat technique of (24), (25) and (27) above for settings listed in table 6. If
oscilloscope calibrator err display does not indicate within limits specified in table 6,
perform b below.
Test instrument
Oscilloscope calibrator
VOLTS/DIV
Vertical deflection CHAN 1 VOLTAGE
err display
switch settings
output
(divisions)
indication (%)
2 mV
5
10 mV
3
5 mV
4
20 mV
3
10 mV
5
50 mV
3
20 mV
5
0.1 V
3
50 mV
4
0.2 V
3
100 mV
5
0.5 V
3
200 mV
5
1.0 V
3
500 mV
4
2.0 V
3
1.0 V
5
5.0 V
3
(29) Reduce all outputs to minimum.
(30) Press corresponding pushbuttons for indications as listed in (a) through (c)
below:
(a) TRIGGER SOURCE to select CH 3.
(b) VERTICAL MODE CH 1 off and CH 3 on.
(c) CH 3 VOLTS/DIV to display .1 V on CRT.
(31) Set oscilloscope calibrator for a CHAN 3, VOLTAGE mode output of 500 mV at
(32) Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to adjust
amplitude for 5 divisions of deflection on TI. If oscilloscope calibrator err display does not
indicate within limits specified in first row of table 7, perform b below.
(33) Repeat technique of (30) (c), (31) and (32) above for settings listed in table 7. If
oscilloscope calibrator err display does not indicate within limits specified in table 7,
perform b below.
Test instrument
Oscilloscope calibrator
VOLTS/DIV
Vertical deflection CHAN 3 VOLTAGE
err display
switch settings
output
(divisions)
indication (%)
0.1 V
5
500 mV
10
0.5 V
4
2V
10
(34) Reduce all outputs to minimum.
(35) Press corresponding pushbuttons for indications as listed in (a) through (c)
below: