TB 9-6625-2295-35
(19) Reduce all outputs to minimum.
(20) Press corresponding pushbuttons for indications as listed in (a) through (f)
below:
(a) VERTICAL MODE CH 1 off and CH 2 on.
(b) TRIGGER SOURCE to CH 2.
(c) TRIGGER COUPLING to AC.
(d) CH 2 switch to 100mV.
(e) X10 MAG to off.
(f) SEC/DIV switch to 50 s.
(21) Set oscilloscope calibrator for a CHAN 2, LEVEL SINE mode output of 50 kHz
and adjust output amplitude for 0.35 division of display on TI.
(22) Adjust TRIGGER LEVEL control for a stable display at all settings listed in
table 13. If a
stable display cannot be
obtained, perform b below.
NOTE
Set A SWP SEC/DIV switch and press X10 MAG pushbutton
as necessary to obtain a well-defined display of test signal.
Table 13. Channel 2
Trigger Level Frequency Response
Stability
Test instrument
Test instrument
trigger coupling
divisions of display
50
kHz
AC
.35
50
MHz
AC
.35
50
MHz
DC
.35
300
MHz
DC
1
300
MHz
AC
1
500
MHz
AC
1.5
500
MHz
DC
1.5
(23) Reduce all outputs to minimum.
a. Performance Check
(1) Press corresponding pushbuttons for indications as listed in (a) through (c) below:
(a) VERTICAL MODE CH1 on and CH2, CH3, and CH 4 off.
(b) TRIGGER MODE to AUTO LVL and TRIGGER SOURCE to VERT CH 1.
(c) Xl0 MAG off.
(2) Set A SWP SEC/DIV switch to 5 ns (knobs locked).
(3) Set oscilloscope calibrator for a CHAN 1, MARKER mode output of 5 ns.
(4) Set CH 1 VOLTS/DIV switch for approximately 2 to 3 divisions of display.
(5) Adjust TRIGGER LEVEL control for a stable display.