TB 9-6625-2295-35
(6) Adjust ⇦POSITION⇨ control to align 2d time marker with 2d vertical graticule
line. Rotate oscilloscope calibrator knob, below EDIT FIELD pushbutton, to align 10th
time marker with 10th vertical graticule line.
If oscilloscope calibrator err readout
indication is not within limits specified in first row of table 14, perform b below.
(7) Repeat technique of (2) through (6) above for remaining settings in table 14. If
oscilloscope calibrator err readout indication is not within limits specified in table 14,
perform b below.
Test instrument
Oscilloscope calibrator
A SWP SEC/DIV
MARKER mode
err display
setting
output
indication (%)
5 ns
5 ns
1.45
10 ns
10 ns
1.45
20 ns
20 ns
1.45
50 ns
50 ns
1.45
100 ns
100 ns
1.45
200 ns
200 ns
1.45
500 ns
500 ns
1.45
1 s
1 s
1.45
2 s
2 s
1.45
5 s
5 s
1.45
10 s
10 s
1.45
20 s
20 s
1.45
50 s
50 s
1.45
100 s
100 s
1.45
200 s
200 s
1.45
500 s
500 s
1.45
1 ms
1 ms
1.45
2 ms
2 ms
1.45
5 ms
5 ms
1.45
10 ms
10 ms
1.45
20 ms
20 ms
1.45
50 ms
50 ms
1.45
100 ms
.1 s
1.45
200 ms
.2 s
1.95
500 ms
.5 s
1.95
500 ms1
50 ms
1.95
Set TI X10 MAG to on.
1
(8) Press Xl0 MAG pushbutton to off.
(9) Set A SWP SEC/DIV switch to 10 ns and B SWP SEC/DIV switch to 5 ns
(knob in) for B SWP only.
(10) Set oscilloscope calibrator for a CHAN 1, MARKER mode output of 5 ns.
(11) Adjust TRIGGER LEVEL control for a stable display.
(12) Adjust ⇦POSITION⇨ control to align 2d time marker with 2d vertical graticule
line. Rotate oscilloscope calibrator knob, below EDIT FIELD pushbutton, to align 10th
15