TB 9-6625-2296-24
NOTE
Assure highlighted area indicates YES before proceeding.
(f) DATA ENTRY ENTER ..........
Factory Defaults restored
(2) Connect measuring receiver sensor module to TI T/R.
(3) Set measuring receiver to measure AM depth, low pass filter 3 kHz, high
pass filter 50 Hz, and PEAK /2 detector selected.
(4) Press key sequence listed in (a) through (h) below:
(a) MODE RF GEN ......................
RF Gen Display
(b) SETUP .....................................
Gen Menu
(c) DATA ENTRY 5 ......................
RF Gen Setup
(d) DATA ENTRY 1 ......................
RF Gen Freq
(e) DATA ENTRY 7 5 0 K ............
750 kHz
(f) DATA ENTRY 2 ......................
RF Gen Level
(g) DATA ENTRY 0 ENTER .......
0 dBm
(h) Ret (F5) ....................................
Return
(5) Record measuring receiver indication as residual AM.
(6) Repeat (4) (e) and (5) above for RF generator frequencies of 10 and 900 MHz.
(7) Press DATA ENTRY 1 2 1 . 1 M keys (121.1 MHz).
(8) Press Source (F4) (Audio Frequency Source), DATA ENTRY 1, and DATA
ENTRY ENTER keys to set source to 1 and AM.
(9) Set measuring receiver to measure modulation distortion, low pass filter 3
kHz, high pass filter 50 Hz and PEAK /2 detector selected.
(10) Press More (F6), A Freq (F1) and press DATA ENTRY keys to set AF FREQ
to 100 Hz, then press DATA ENTRY ENTER key.
(11) Press More (F6) key twice, Mod (F5) and DATA ENTRY keys to set
MODULATION to 50%, then press DATA ENTRY ENTER keys.
(12) Measuring receiver distortion indication will be less than specified in table 16
for current TI setting.
(13) Repeat (7) and (9) through (12) for remaining settings listed in table 16.
53