Figure 7. Power Meter Setup (20 mW 500 mW Range).
Table 15. Power Meter Accuracy (1 W 200 W).

Calibration Procedure For Radio Test Set An/Grm-114B, An/Grm-122, And Ts-4317/Grm Page Navigation
  43    44    45    46    47  48  49    50    51    52    53  
img
. . .
TB 9-6625-2296-24
Table 14. Power Meter Accuracy (20 mW 500 mW).
Test
Source
Source
Measuring
Test instrument indications
instrument
output
output
receiver
RANGE
amplitude
frequency
indications
Min
Max
20 mW
10.00
mW
10  MHz
10.00 mW
8.9 mW
11.1 mW
50 mW
25.00
mW
10  MHz
25.00 mW
22.4 mW
27.6 mW
100 mW
50.00
mW
10  MHz
50.00 mW
44
mW
56  mW
200 mW
100.0
mW
10  MHz
100.0  mW
89
mW
111  mW
500 mW
250.0
mW
10  MHz
250.0  mW
224
mW
276  mW
20 mW
10.00
mW
400  MHz
10.00 mW
8.9 mW
11.1 mW
50 mW
25.00
mW
400  MHz
25.00 mW
22.4 mW
27.6 mW
100 mW
50.00
mW
400  MHz
50.00 mW
44
mW
56  mW
200 mW
100.0
mW
400  MHz
100.0  mW
89
mW
111  mW
500 mW
250.0
mW
400  MHz
250.0  mW
224
mW
276  mW
20 mW
10.00
mW
750  MHz
10.00 mW
7.9 mW
12.1 mW
50 mW
25.00
mW
750  MHz
25.00 mW
19.9 mW
30.1 mW
100 mW
50.00
mW
750  MHz
50.00 mW
39
mW
61  mW
200 mW
100.0
mW
750  MHz
100.0  mW
79
mW
121  mW
500 mW
250.0
mW
750  MHz
250.0  mW
199
mW
301  mW
(8) Reduce signal generator and RF power amplifier gain to minimum and
disconnect equipment setup.
(9) Press Range (F1) and DATA SCROLL arrow keys to select 20 mW range and
press Zero (F4) key. Press DATA ENTRY ENTER key when Zero submenu appears.
Adjust DATA SCROLL knob to select 1 W range and press the DATA ENTRY ENTER key.
(10) Connect equipment as shown in figure 8 below.
NOTE
Use directional power sensor model 4021.
Figure 8. Power Meter Setup (1 W 200 W Range).