TB 9-6625-2296-24
(a) DATA SCROLL arrow keys ...
Highlight DEV
(b) DATA ENTRY ENTER ..........
Select DEV
(c) Range (F1) ...............................
Deviation Range
(d) DATA SCROLL arrow keys ...
Highlight 20 kHz range
(e) DATA ENTRY ENTER ..........
Select 20 kHz Deviation
(f) Ret (F6) ....................................
Return
(g) FM Z (F3) .................................
FM Zero
(15) Press MODE MTRS (Meters Menu), AUX (F6) (Auxiliary Functions Menu), and
DATA ENTRY ENTER keys; press FIELD SELECT arrow keys to highlight 2. Dev
Meter (Peak) and press DATA ENTRY ENTER key.
(16) Press DATA ENTRY ENTER and DATA ENTRY keys to enter value
displayed on measuring receiver in 20 kHz (+) data field, and press DATA ENTRY
ENTER key.
(17) Press DATA ENTRY - key, followed by DATA ENTRY numeric keys to enter
value displayed on measuring receiver in the 20 kHz(-) data field, and press DATA ENTRY
ENTER key.
(18) Press Ret (F5) (Return), MODE RCVR (Receiver Display), SHIFT, and Y keys
(Backup Cal Data ? y/n).
(19) Adjust signal generator output for 121.1 MHz at 0 dBm with 45 kHz deviation at
3 kHz rate and set FM off.
(20) Press key sequence listed in (a) through (g) below:
(a) DATA SCROLL arrow keys ...
Highlight DEV
(b) DATA ENTRY ENTER ..........
Select DEV
Deviation Range
(c) Range (F1) ...............................
Highlight 50 kHz range
(d) DATA SCROLL arrow keys ...
(e) DATA ENTRY ENTER ..........
Select 50 kHz Deviation
(f) Ret (F6) ....................................
Return
(g) FM Z (F3) .................................
FM Zero
(22) Press MODE MTRS (Meters Menu), AUX (F6) (Auxiliary Functions Menu), and
DATA ENTRY ENTER keys; press FIELD SELECT arrow keys to highlight 2. Dev
Meter (Peak) and press DATA ENTRY ENTER key.
(23) Press DATA ENTRY ENTER key and press DATA ENTRY keys to enter value
displayed on measuring receiver in 50 kHz(+) data field. Press DATA ENTRY ENTER key.
66