TB 9-6625-2296-24
(24) Press DATA ENTRY - key, followed by DATA ENTRY numeric keys to enter
value displayed on measuring receiver in 50 kHz(-) data field, and press DATA ENTRY
ENTER key.
(25) Adjust signal generator output for 121.1 MHz at 0 dBm with 75 kHz deviation at
3 kHz rate and set FM off.
(26) Press Ret (F5) (Return), MODE RCVR (Receiver Display), SHIFT, and Y keys
(Backup Cal Data ? y/n).
(27) Press key sequence listed in (a) through (g) below:
Highlight DEV
(a) DATA SCROLL arrow keys ...
(b) DATA ENTRY ENTER ..........
Select DEV
(c) Range (F1) ...............................
Deviation Range
(d) DATA SCROLL arrow keys ...
Highlight 100 kHz range
(e) DATA ENTRY ENTER ..........
100 kHz deviation
(f) Ret (F6) ....................................
Return
(g) FM Z (F3) .................................
FM Zero
(29) Press MODE MTRS (Meters Menu), AUX (F6) (Auxiliary Functions Menu), and
DATA ENTRY ENTER keys; press FIELD SELECT arrow keys to highlight 2. Dev
Meter (Peak) and press DATA ENTRY ENTER key.
(30) Press DATA ENTRY ENTER and DATA ENTRY keys to enter value
displayed on measuring receiver in 100 kHz(+) data field. Press DATA ENTRY ENTER
key.
(31) Press DATA ENTRY - key, followed by DATA ENTRY numeric keys to enter
value displayed on measuring receiver in 100 kHz(-) data field, and press DATA ENTRY
ENTER key.
(32) Press Ret (F5) (Return), MODE RCVR (Receiver Display), SHIFT, and Y keys
(Backup Cal Data ? y/n).
23. Amplitude Modulation Meter
a. Performance Check
(1) Press key sequence listed in (a) through (f) below:
(a) MODE MTRS ..........................
Meters Menu
(b) AUX (F6) ..................................
Auxiliary
(c) MEMORY RCL .......................
Recall
(d) FIELD SELECT ..................
Highlight 10. Factory Defaults
(e) DATA ENTRY ENTER ..........
Factory Defaults recall