TB 9-6625-2296-35
(2) Press MODE RCVR (Receiver Display), Freq (F1) (Frequency), DATA ENTRY
1, and M (1 MHz), and (F3) T/R keys.
(3) Connect frequency counter 10 MHZ OUT to function/arbitrary generator 10
MHz In.
(4) Connect function/arbitrary generator OUTPUT to TI T/R and frequency counter
A input.
(5) Press MODE MTRS key (Meters Menu) and DATA ENTRY 2 key (Freq Error
Meter/Freq Meter).
(6) Set function/arbitrary generator for a 1 MHz, 1 V p-p output.
(7) Adjust function/arbitrary generator frequency until TI FREQ (MHz) display
indicates 1.000000 MHz.
(8) Frequency counter will indicate 1 MHz 0.5 Hz. TI FREQ ERR display will
indicate within 0.5 Hz.
(9) Adjust function/arbitrary generator frequency until TI FREQ (MHz) display
indicates 1.000020 MHz.
(10) Frequency counter will indicate 1.000020 MHz 0.5 Hz. TI FREQ ERR display
will indicate between +19.5 and +20.5 Hz.
(11) Adjust function/arbitrary generator frequency until TI FREQ (MHz) display
indicates 0.999980 MHz.
(12) Frequency counter will indicate .999980 MHz 0.5 Hz. TI FREQ ERR display
will indicate between -19.5 and -20.5 Hz.
(13) Disconnect function/arbitrary generator from frequency counter and TI.
(14) Connect frequency counter 10 MHz OUT to signal generator EXT REF INPUT.
(15) Connect signal generator RF OUTPUT to frequency counter C input and to
TI T/R.
(16) Press key sequence listed in (a) through (n) below:
Receiver Display
(a)
MODE RCVR ..........................
Setup Screen
(b)
SETUP .....................................
Set Rcvr Freq
(c)
DATA ENTRY 1 ......................
900
(d)
DATA ENTRY 9 0 0................
900 MHz
(e)
DATA ENTRY ENTER ..........
Select Mod
(f)
DATA ENTRY 2 ......................
(g)
FIELD SELECT arrow keys ..
Highlight 11. User Defined
(h)
DATA ENTRY ENTER ..........
Select 11. User Defined
IF Filters
(i)
DATA ENTRY 2 ......................
300 kHz
(j)
DATA ENTRY 3 ......................
Rcvr Menu
(k)
F6 (Esc) ....................................
Receiver Display
(l)
F5 (Ret) ....................................