TB 9-6625-2296-35
NOTE
Random numbers may be displayed as test is running. The
testing process can be halted by pressing the STOP key
anytime during the testing process. If a test fails, the
Diagnostic Menu appears. The operator can view the Test
Results of the failed test (Disp (F3)), perform a Debug
Operation on the failed test (Debug (F2)), retest failed test
(Cont (F1)) or retest from the beginning of the testing process
(Restart (F6)).
Return to SINCGARS Verification Test
(m) Rtn (F6) ...................................
Menu
NOTE
Attach Test Fixture TF-119A to the UUT Connector on the PL-
1536/GRM.
BIT Level 2 Test
(n) BIT 2 (F3) ................................
BIT Level 2 Test begins
(o) GO .............................................
NOTE
If a test fails, the Diagnostic Menu appears. The operator can
view the Test Results of the failed test (Disp (F3)), perform a
Debug Operation on the failed test (Debug (F2)), retest failed
test (Cont (F1)) or retest from the beginning of the testing
process (Restart (F6)).
(2) Remove Test Fixture TF-119A and proceed to paragraph 29 Final Procedure.
b. Adjustments. No adjustments can be made. For test failures refer to TM 11-6625-3246-40.
28. J-4843/GRM Self Test (AN/GRM-114B only)
a. Performance Check
(1) Press key sequence listed in (a) through (m) below:
Meters Menu
(a)
MODE MTRS ..........................
Auxiliary
(b)
AUX (F6) ..................................
Recall
(c)
MEMORY RCL .......................
FIELD SELECT ⇑ ..................
Highlight 10. Factory Defaults
(d)
Factory Defaults recall
(e) DATA ENTRY ENTER............