TB 9-6625-2303-24
20. Plug-in Element 4410-4 Calibration
a. Performance Check
(1) Connect equipment as shown in figure 2 above.
(2) Insert plug-in element into TI and rotate until the arrow points in the direction
of the 50 Ω termination.
(3) Set TI range switch to 1 and allow equipment to warm-up for 5 minutes.
(4) Adjust signal generator and RF power amplifier frequency controls for 30 MHz.
(5) Adjust signal generator and RF power amplifier amplitude controls for a TI
indication of 100 W. RF power meter will indicate within limits in table 15.
(6) Repeat technique of steps (4) and (5) above for remaining settings listed in
table 15. RF power meter will indicate within limits in table 15.
Table 15. 4410-4 Element Accuracy.
Test instrument
Test instrument
RF power meter indication (W)
Signal generator
Low pass
range switch
full scale reading
frequency (Hz)
filter used
Min
Max
settings
(W)
1001
100001
95001
105001
301
30001
28501
31501
101
10001
9501
10501
31
3001
2851
3151
30 M
TLC30-4EF7
1
100
95
105
0.3
30
28.5
31.5
0.1
10
9.5
10.5
0.1
10
9.5
10.5
0.3
30
28.5
31.5
1
100
95
105
31
3001
2851
3151
16 M
TLC20-6EF1
101
10001
9501
10501
301
30001
28501
31501
1001
100001
95001
105001
1001
100001
95001
105001
301
30001
28501
31501
101
10001
9501
10501
31
3001
2851
3151
2M
1-60-561-3
1
100
95
105
0.3
30
28.5
31.5
0.1
10
9.5
10.5
Not calibrated due to standards limitations.
1
(7) Reduce outputs to minimum and remove plug-in element from TI.
b. Adjustments. No adjustments can be made.