• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Plug-in Element 4410-4 Calibration
Plug-in Element 4410-2 Calibration

Calibration Procedure For Radio Frequency 4410 Series Wattmeter And Plug-In Elements Page Navigation
  14    15    16    17    18  19  20    21    22  
img
. . . . . . .
TB 9-6625-2303-24
21. Plug-in Element 4410-3 Calibration
a. Performance Check
(1) Connect equipment as shown in figure 2 above.
(2) Insert plug-in element into TI and rotate until the arrow points in the direction
of the 50 Ω termination.
(3) Set TI range switch to 1 and allow equipment to warm-up for 5 minutes.
(4) Adjust signal generator and RF power amplifier frequency controls for 30 MHz.
(5) Adjust signal generator and RF power amplifier amplitude controls for a TI
indication of 100 W. RF power meter will indicate within limits in table 16.
(6) Repeat technique of steps (4) and (5) above for remaining settings listed in
table 16. RF power meter will indicate within limits in table 16.
Table 16. 4410-3 Element Accuracy.
Test instrument
Test instrument
RF power meter indication (W)
Signal generator
Low pass
range switch
full scale reading
frequency (Hz)
filter used
Min
Max
settings
(W)
1001
10001
9501
10501
301
3001
2851
3151
10
100
95
105
3
30
28.5
31.5
30 M
TLC30-4EF7
1
10
9.5
10.5
0.3
3
2.85
3.15
0.1
1
0.95
1.05
0.1
1
0.95
1.05
0.3
3
2.85
3.15
1
10
9.5
10.5
16 M
TLC20-6EF1
3
30
28.5
31.5
10
100
95
105
301
3001
2851
3151
1001
10001
9501
10501
1001
10001
9501
10501
301
3001
2851
3151
10
100
95
105
3
30
28.5
31.5
2M
1-60-561-3
1
10
9.5
10.5
0.3
3
2.85
3.15
0.1
1
0.95
1.05
Not calibrated due to standards limitations.
1
(7) Reduce outputs to minimum and remove plug-in element from TI.
b. Adjustments. No adjustments can be made.
20


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.