TB 9-6625-2304-35
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 11-6625-3267-24&P.
d. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Open protective cover from TI when necessary to make adjustments and as
instructed in text. Close cover after completing the adjustments.
b. Connect calibrator and dc
power supply as
shown in
figure 1.
c. Set calibrator for a 28 V dc and adjust dc power supply for a 36 V dc output.
Periodically monitor output of dc power supply to maintain the 36 (1.0) V dc.
d. Allow 1 hour for equipment warm-up before proceeding.