TB 9-6625-2304-35
a. Performance Check
(1) Configure multimeter for most accurate dc voltage measurements.
(2) Connect multimeter INPUT HI to TI TP12 and LO to TP23. Multimeter will
indicate between 0.089550 and 0.090450 V dc.
(3) Repeat technique of (2) above while moving multimeter INPUT HI test
lead to each TI test point (TP) specified in table 3. Multimeter will indicate within