Equipment Setup
Figure 3. Horizontal scale error - equipment setup.

Calibration Procedure For Optical Fiber Test Set, Ts-4320(P)/G Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . .
TB 9-6625-2309-24
1....
Figure 1. Example of correct leading edge position of dead zone pulse (TD-260C).
(5) If leading edge is not visible or is located to right of the origin (dashed lines,
figure 2), set ON-OFF switch to OFF and repeat steps (1) through (4) above.
Figure 2. Example: Example of incorrect leading edge position of dead zone pulse (TD-260C).
c. Connect equipment as shown in figure 3, Connection A.  Connect optical fiber
jumper cable from FOCUS-SWCM OPTICAL I-O to ST connector adapter attached to
optical input of LIGHTWAVE MULTIMETER (fig. 3).