TB 9-6625-2311-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin provides instructions for the
was used as the prime data source in compiling these instructions. The equipment
being calibrated will be referred to as the TI (test instrument) throughout this
bulletin.
a. Model Variations. None
b. Time and Technique.
The time required for this calibration is
approximately 4 hours, using the dc and low frequency and microwave techniques.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels
are prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in
which they appear. When adjustments are in tables, the (R) follows the designated
adjustment. Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
Performance specifications
parameters
Boonton
Accuracy: +(% of FS + % of reading)
Model 9200B
Voltage
Frequency range
Tolerance
reading+ 3 V
Range 1
1 MHz
3%
0.200 to 1.000 mV
10
kHz to 100 MHz
4%
reading+ 3 mV
100
MHz to
1 GHz
6%
reading+ 3 mV
1
GHz to
1.2 GHz
10%
reading+ 3 mV
Range 2
1
MHz
2%
reading+ 2 mV
1.000 to 3.000 mV
10
kHz to 100 MHz
3%
reading+ 2 mV
100
MHz to
1 GHz
5%
reading+ 2 mV
1
GHz to
1.2 GHz
9%
reading+ 2 mV
Range 3
1
MHz
1%
reading+ 10 mV
3.00
to 30.00
mV
10
kHz to 100 MHz
2%
reading+ 10 mV
100
MHz to
1 GHz
4%
reading+ 10 mV
1
GHz to
1.2 GHz
8%
reading+ 10 mV