Figure 1. Frequency response - equipment setup.
Figure 2. Boonton, Model 9200B switch S1 detail.

Calibration Procedure For Rf Millivoltmeter Boonton, Model 9200B Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . . . . . .
TB 9-6625-2311-24
(22) Set up measuring receiver to measure RF power in mV.
(23) Adjust signal generator frequency controls to frequency settings listed in
table 3 below.
(24) Adjust signal generator amplitude controls for a 100.0 mV indication on
measuring receiver. Record TI indication under TI instrument indications No. 1
column in table 3.
(25) Repeat technique of (23) and (24) above for remaining signal generator
frequency settings listed in table 3 below.
(26) Set signal generator output to off.
(27) Reverse the output port connections of the power splitter.
(28) Set signal generator output to on.
(29) Repeat technique of (23) and (24) above and record TI indication under TI
indications No. 2 column in table 3 below.
NOTE
Refer to table 3 for (30) below.
(30) Add TI indications No. 1 value to TI indications No. 2 value and divide
the sum by two. Record results in TI indications average column. Results recorded
in TI indications average column will be within the limits specified in the minimum-
maximum columns of table 3.
(31) Repeat (30) above for each row listed in table 3.
Table 3. Frequency Response
Test instrument indications
Signal generator
frequency control
settings
No. 1
No. 2
Average
Min
Max
(MHz)
(mV)
(mV)
(mV)
(mV)
(mV)
10.00
97.9
102.1
30.00
97.9
102.1
60.00
97.9
102.1
80.00
97.9
102.1
90.00
97.9
102.1
110.00
95.9
104.1
200.00
95.9
104.1
400.00
95.9
104.1
600.00
95.9
104.1
800.00
95.9
104.1
900.00
95.9
104.1
1100.00
91.9
108.1
1200.00
91.9
108.1