Table 3. Frequency Response
RF Millivoltmeter Accuracy

Calibration Procedure For Rf Millivoltmeter Boonton, Model 9200B Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. .
TB 9-6625-2311-24
b. Adjustments
(1) Set calibrator to standby.
(2) Set internal switch S1 bit 2 in the open position (fig. 2).
Figure 2. Boonton, Model 9200B switch S1 detail.
(3) Perform (a) through (f ) below with the TI keys:
(a) Press 0.
(b) Press Z0.
(c) Press 0.
(d) Press MODE dBV.
(e) Press MODE mV.
(f) Press ZERO.
(4) Wait for zeroing process to complete. If the zero indication is not less
than +50 counts, repeat (3) (f ) above until this is achieved.
(5) Adjust calibrator controls for an output of 0.7 V at 10 kHz.
(6) Perform (a) through (e) below with the TI keys:
(a) Press 0.
(b) Press MODE dBV.
(c) Press 7, . , 0, 0, MODE dBmV and wait for a stable reading.
(d) Press AVERAGE SEL.
8