Performance Check
Table 5. Resistance.

Calibration Procedure For Precision Component Analyzer, Model 6425B Page Navigation
  1    2    3    4    5  6  7  
img
. . . . . . . . .
TB 9-6625-2313-40
11. Inductance
NOTE
It is important that trimming operations are made with Kelvin
leads arranged as they will be for measurement checks.
Measurements may be affected by movement of the leads. If an
out-of-tolerance condition exists, repeat the trimming operation
and the measurement check.
a. Performance Check
(1) Repeat technique of paragraph 9 above.
(2) Connect appropriate inductance standard from table 4 to TI.
(3) Press soft keys for L, R, SERIES, and REP settings.
(4) The TI inductance indication will be within 0.10% of standard's test report
value; if not, perform b below.
Table 4. Inductance Measurement.
Inductance standard
(nominal value)
100
H
1
mH
10
mH
100
mH
1
H
10
H
(5) Press soft key for Single trigger setting.
(6) Repeat technique of (2) through (4) above for the remaining values in table 4.
b. Adjustments. Repeat paragraphs 9 and 11 above being careful with movement of
Kelvin leads. Otherwise, no adjustments can be made.
12. Resistance
NOTE
It is important that trimming operations are made with Kelvin
leads arranged as they will be for measurement checks.
Measurements may be affected by movement of the leads. If an
out-of-tolerance condition exists, repeat the trimming operation
and the measurement check.
a. Performance Check
(1) Repeat technique of paragraph 9 above.
(2) Connect resistance standard No. 1 to TI.
7