Table 4. Inductance Measurement.

Calibration Procedure For Precision Component Analyzer, Model 6425B Page Navigation
  2    3    4    5    6  7
img
. . . . . .
TB 9-6625-2313-40
(3) For resistance below 1000 ohms, press soft keys L, R, and Series; for resistance
above 1000 ohms press soft keys C, R and PARALLEL.
(4) Set resistance standard to the first value in table 5.
(5) The TI's resistance indication will be within 0.05% of the resistance standard
value; if not, perform b below.
(6) Repeat steps 3 and 4 for the remaining values in table 5.
Table 5. Resistance.
Resistance (kΩ)
0.005
0.10
1
10
100
500
1000
5000 1
10000
1.
Use resistance standard No. 2.
b. Adjustments. Repeat paragraphs 9 and 12 above being careful with movement of
Kelvin leads. Otherwise, no adjustments can be made.
13. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.