TB 9-6625-2313-50
(1) 10.0 kHz.
(2) 1.00 Vac.
(3) 0.0 Vdc.
(4) Bias Off.
(5) C.
(6) G.
(7) PARALLEL.
(8) Auto.
(9) REP.
(10) SLOW.
e. From the keypad press CODE, 9, then ENTER keys. The upper left of crt display
should read *8.
f. Place the Kelvin leads in a position that will approximate the measurement position.
Connect leads as illustrated in figure 1 for TRIM O/C. Ensure that the clips are separated
by at least 5 cm (approximately 2.5 inches).
NOTE: MAY ALSO CLIP LEAD(S) TO A PIECE OF BARE WIRE.
g. The
residual capacitance reading should not
exceed 8.8 pF, with
parallel
h. Press TRIM O/C pushbutton, (allow 10-15 seconds) then press TRIGGER
pushbutton to initiate auto-trim process. The display should read 0.0 fF (2.0 fF) and 0.0 nS
(0.04 ns).
i. Connect leads as
illustrated in
figure 1
for TRIM
S/C, ensuring that the
lead clips
are not touching each other. The analyzer should autorange to range *1 and the drive
level should change to 100 mA.
5