• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
SECTION I. IDENTIFICATION AND DESCRIPTION
SECTION II. EQUIPMENT REQUIREMENTS

Calibration Procedure For Multimeter, Model 34401A Page Navigation
  1  2  3    4    5    6    7  
img
TB 9-6625-2315-24
Table 1. Calibration Description - Continued
Test instrument
parameters
Performance specifications
Dc Current
Range:  10 mA to 3 A
Accuracy:  (% of reading + % of range) 6 digits, 1 hour warm-up
Range
Accuracy
10.00000
mA
0.050 + 0.020
100.0000
mA
0.050 + 0.005
1.000000  A
0.100 + 0.010
3.000000  A
0.120 + 0.020
Ac Voltage
Range:  100 mV to 750 V
Accuracy:  (% of reading + % of range) 6 digits, 1 hour warm-up,
slow ac filter, sinewave inputs > 5 % of range
Range
Frequency
Accuracy
100.0000
mV
3 Hz
-
5 Hz
1.00 + 0.04
5 Hz
-  10 Hz
0.35 + 0.04
10 Hz
-  20 kHz
0.06 + 0.04
20 kHz -  50 kHz
0.12 + 0.05
50 kHz - 100 kHz
0.60 + 0.08
100 kHz - 300 kHz
4.00 + 0.50
1.000000 V
3
Hz
-
5
Hz
1.00 + 0.03
to
5
Hz
-  10
Hz
0.35 + 0.03
750.000
V
10
Hz
-  20
kHz
0.06 + 0.03
20
kHz
-  50
kHz
0.12 + 0.05
50
kHz
- 100
kHz
0.60+ 0.08
100
kHz
- 300
kHz
4.00 + 0.50
Ac Current
Range:  1 A and 3 A
Frequency: 3 Hz to 300 kHz
Accuracy:  (% of reading + % of range) 6 digits, 1 hour warm-up,
slow ac filter, sinewave inputs > 5 % of range
Range
Frequency
Accuracy
1.000000  A
3 Hz
-
5 Hz
1.00 + 0.04
5 Hz
-  10 Hz
0.30 + 0.04
10 Hz
-
5 Hz
0.10 + 0.04
3.000000
A
3 Hz
-
5 Hz
1.10 + 0.06
5 Hz
-
10 Hz
0.35 + 0.06
10 Hz
-
5 kHz
0.15 + 0.06
Frequency
Range:  100 mV to 750 V
Frequency: 3.0 Hz to 300 kHz
Accuracy:
(% of reading) 6 digits, 1 hour warm-up
3 Hz
-
5 Hz
0.10
5 Hz
-  10 Hz
0.05
10 Hz
-  40 Hz
0.03
40 Hz
- 300 kHz  0.01
1Specifications
are for 4-wire ohms function, or 2-wire ohms using Math Null.
3


Privacy Statement