Resistance
Ac Current

Calibration Procedure For Multimeter, Model 34401A Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. . . . . . . . .
TB 9-6625-2315-24
(13) Set calibrator EX SNS off and output to 100 M . Using output adjustment
controls set calibrator control display Reading equal to TI indication. Calibrator control
display Error will indicate within 0.81 percent.
b. Adjustments. Perform paragraphs 14 and 15 below.
10. Dc Current
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI front Input I and LO.
(2) Press Shift key then FUNCTION DC V/DC I key.
(3) Press Shift key then RANGE/DIGITS Auto/Man key to select 6 digits resolution.
(4) Set calibrator for a 0 mA dc output.
(5) Press RANGE/DIGITS  or  key to select ranges listed in table 7. If TI does
not indicate within limits specified, perform b below.
Table 7. Dc Current Zero Offset
Test instrument
Indications
Ranges
Min
Max
10 mA
-00.002  mA
+00.002  mA
100 mA
-000.005  mA
+000.005  mA
1A
-0.0001 A
+0.0001 A
3A
-0.0006 A
+0.0006 A
(6) Press RANGE/DIGITS  or  key to select TI ranges and set calibrator output
for settings listed in table 8. If TI does not indicate within limits specified, perform b below.
Table 8. Dc Current
Test instrument
Calibrator
Test instrument indications
ranges
output settings
Min
Max
10
mA
10
mA
9.993
mA
10.007
mA
100
mA
100
mA
99.945
mA
100.055
mA
1
A
1
A
0.9989
A
1.0011
A
3
A
2
A
1.997
A
2.003
A
b. Adjustments. Perform paragraphs 14 and 15 below.
11. Ac voltage
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI Input HI and LO.
(2) Press keys as listed in (a) through (f) below:
(a) FUNCTION AC V.
8