TB 9-6625-2315-24
Test instrument
Calibrator
Value displayed in
ranges
output
(10) above
Voltage
100 mV
100 mV
50 kHz
100.000
mV AC
1V
1
V
1 kHz
1.00000 V AC
10 V
10
V
1 kHz
10.0000
V AC
10 V
10
V
50 kHz
10.0000
V AC
10 V
10
V
10 Hz
10.0000
V AC
100 V
100
V
1 kHz
100.000
V AC
100 V
100
V
50 kHz
100.000
V AC
750 V
750
V
1 kHz
750.00
V AC
750 V
200
V
50 kHz
200.00
V AC
(13) Press calibrator RESET key and disconnect from TI.
f. Perform frequency adjustments as listed in (1) through (12) below:
(1) Connect function/arbitrary generator Output to TI Input HI and LO using
50
feedthrough termination.
(2) Press FUNCTION Freq key.
or
to select 1 V range.
(3) Press RANGE/DIGITS
(4) Set function/arbitrary generator for a sine wave 50 kHz, 1 Vrms output.
key.
(5) Press Shift key then MENU
or
key until F: CAL MENU is displayed.
(6) Press MENU
(7) Press RANGE/DIGITS
key until 1: UNSECURED is displayed.
(8) Press MENU
key until 2: CALIBRATE is displayed.
key until PARAMETERS is displayed.
(9) Press RANGE/DIGITS
NOTE
The calibration parameter initial display will be 0.000,000 HZ.
HZ will have to be changed to 00.000,00 kHz in order to enter
50.000,00 kHz in (10) below.
or
and RANGE/DIGITS
or
keys until 50.000,00 kHz
(10) Press MENU
is displayed.
(11) Press RANGE/DIGITS Auto/Man key. Adjustment takes approximately 20
seconds to complete then TI will return to measurement mode.
(12) Press function/arbitrary generator Output key to off and disconnect from TI.
g. Press TI keys as listed in (1) through (6) below:
to display A: MEAS MENU.
(1) Shift then MENU
(2) MENU
until F: CAL MENU is displayed.
until 1: UNSECURED is displayed.
(3) RANGE/DIGITS
14