TB 9-6625-2317-40
Table 1. Calibration Description Continued.
Test instrument parameters
Performance specifications
Phase modulation
Range: 10 MHz to 1.8 GHz
Rate:
300 Hz to 4 kHz
Accuracy: 3% rdg +3 counts
Demodulation range: 30 Hz to 100 kHz
Accuracy: Frequency standard accuracy +1 count
Range: +10 to +30 dBm4
Accuracy: 3 dB
1 30 minute warm-up.
2 0.3 ppm for 24 hour period after a 2 hour warm-up at a constant ambient temperature.
3 Not verified with this procedure.
4 +20 dBm (100 mW) maximum without pad. +30 dBm (1 W) maximum with 10 dB pad.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Reference Calibration
Standards Set, NSN 4931-00-621-7878. Alternate items may be used by the calibrating
activity. The items selected must be verified to perform satisfactorily prior to use and must
bear evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and TI. Where the four-to-one ratio cannot be met, the actual
accuracy of the equipment selected is shown in parentheses.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in the
calibration procedure. The following peculiar accessory is also required for this calibration:
RF Power Amplifier, Ophir RF, Model XRF733 and Low Pass Filters, Telonic Series.
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
AUDIO ANALYZER
Source range: 1 kHz
Boonton, Model 1121
Range: 153 mV to 764 mV
(1121)
Analyzer distortion: <0.15%
CALIBRATION
No substitute
Agilent, Model 11715A
FIXTURE
(11715A)
CALIBRATOR
Range: 1 kHz to 4 kHz
Fluke, Model 5720A (5720A)
(p/o MIS-35947); w/
Accuracy: 0.025%
amplifier, Fluke 5725A/AR
(5725A/AR)
FREQUENCY
Range: 1 kHz to 1.5 GHz
Fluke, Model PM6681/656
COUNTER
(PM6681/656)
Accuracy: 0.075 ppm
POWER METER
Range: 160 kHz to 1.6 GHz
Agilent, Model E4419B
Range: 0 to -25 dBm
(E4419B) w/power sensor
Agilent, Model E9304AH18
Accuracy: 0.25 dB
(E9304AH18)