TB 9-6625-2317-50
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Phase modulation
Range: 10 MHz to 1.8 GHz
Rate: 300 Hz to 4 kHz
Accuracy: 3% of reading +3 counts
Demodulation range: 30 Hz to 100 kHz
Accuracy: Frequency standard accuracy + one count
Power measurement
Range: +10 to +30 dBm 4
Accuracy: 3 dB
130
minute warm-up.
20.3
ppm for 24 hour period after a 2 hour warm-up at a constant ambient temperature.
3Not
verified with this procedure.
4+20
dBm (100 mW) maximum without pad. +30 dBm (1 W) maximum with 10 dB pad.
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Reference Calibration
Standards Set, NSN 4931-00-621-7878. Alternate items may be used by the calibrating
activity. The items selected must be verified to perform satisfactorily prior to use and must
bear evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and TI. Where the four-to-one ratio cannot be met, the actual
accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
AUDIO ANALYZER
Source range: 1 kHz
Boonton, Model 1121
Range: 153 mV to 764 mV
(1121)
Analyzer distortion: <0.15%
CALIBRATION
No substitute
Hewlett-Packard Model
FIXTURE
11715A (11715A)
CALIBRATOR
Range: 1 kHz to 30 MHz
Fluke, Model 5720A (5700A/EP)
(p/o MIS-35947; w/ac divider,
Accuracy: 0.025%
Fluke, Model 7405A-4207
Range: 114.5 mV to 890 mV (+12 dBm)
(7405A-4207)
Accuracy: 0.75 dBm
FREQUENCY
Range: 1 kHz to 1.5 GHz
Fluke, Model PM6681/656
COUNTER
(PM6681/656)
Accuracy: 0.075 ppm
MEASURING
Range:
12.5 MHz to 100 MHz
Hewlett-Packard, Model
RECEIVER
Rate:1000 Hz
8902A w/sensor, HP Model
Deviation:
30 to 100 kHz
11722A (11722A)
Accuracy: .75% (1%)
Depth:
40% to 80%
Accuracy: .625% (1%)
Deviation:
Accuracy: .75% (1%)