• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
SECTION III. CALIBRATION PROCESS
Figure 2. Frequency resolution accuracy hook-up.

Calibration Procedure For Signal Generator Model 68369Nv Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . .
TB 9-6625-2323-24
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(s)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Remove TI protective cover only as necessary to make adjustments.
CAUTION
Before connecting TI to power source, make sure TI is set to the
power source line voltage as shown on rear of TI.
b. Connect TI to 115 V ac power source.
c. Press LINE switch to OPERATE and allow at least one hour for TI to warm-up and stabilize.
d. Connect equipment as shown in figure 1.
Figure 1. Time base verification hook-up.
NOTE
TI must have power applied and in STANDBY or OPERATE
for at least 120 continuous hours before the frequency drift
specification can be verified.
e. Verify that the frequency difference meter drift is less than 5 parts in 10-8 per day.
f. Disconnect frequency difference meter from the TI and the time/frequency workstation.
8. Frequency Accuracy
a. Performance Check
(1) Connect equipment as indicated in figure 2.
5


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.