High Frequency Response
Low and Midband Accuracy - Continued

Calibration Procedure For True Rms Voltmeter, Models 8922A And 8922A/Aa Page Navigation
  1    2    3    4    5  6  7    8    9  
TB 9-6625-2324-24
(6) Set calibrator for a 1 V, 1 kHz output. TI will indicate between 0.995 and 1.005 V.
Record TI indication.
(7) Set calibrator for a 3 V, 1 kHz output. TI will indicate between 2.985 and 3.015 V.
Record TI indication.
(8) Connect calibrator WIDEBAND output to TI INPUT using calibrator wideband
cable and 50  feedthrough termination.
(9) Set calibrator for a 1 mV, 1 kHz wideband output.
(10) Adjust calibrator amplitude output adjustment controls until TI indication
equals TI indication recorded in (3) above, then press calibrator NEW REF key.
(11) Set calibrator to 10 MHz and adjust amplitude output adjustment controls until
TI indication equals indication recorded in (3) above. If calibrator Error display indication
is not within 5 percent, perform b (1) below.
(12) Set calibrator to 1.5 MHz and adjust amplitude output adjustment controls until
TI indication equals indication recorded in (3) above. Calibrator Error display indication
will be within 4 percent.
(13) Repeat (12) above at 990 kHz and 190 kHz.
(14) Set calibrator for a 10 mV, 1 kHz wideband output.
(15) Adjust calibrator amplitude output adjustment controls until TI indication
equals TI indication recorded in (4) above, then press calibrator NEW REF key.
(16) Set calibrator to 10 MHz and adjust amplitude output adjustment controls until
TI indication equals TI indication recorded in (4) above.  If calibrator Error display
indication is not within 5 percent, perform b (2) and (3) below.
(17) Set calibrator to 1.5 MHz and adjust amplitude output adjustment controls until
TI indication equals TI indication recorded in (4) above.  Calibrator Error display
indication will be within 4 percent.
(18) Set calibrator to 990 kHz and adjust amplitude output adjustment controls until
TI indication equals TI indication recorded in (4) above.  Calibrator Error display
indication will be within 2 percent.
(19) Set calibrator to 190 kHz and adjust amplitude output adjustment controls until
TI indication equals TI indication recorded in (4) above.  Calibrator Error display
indication will be within 1 percent.
(20) Set calibrator for a 100 mV, 1 kHz wideband output.
(21) Adjust calibrator amplitude output adjustment controls until TI indication
equals TI indication recorded in (5) above, then press calibrator NEW REF key.
(22) Set calibrator to 190 kHz and adjust amplitude output adjustment controls until
TI indication equals TI indication recorded in (5) above.  Calibrator Error display
indication will be within 0.5 percent.
(23) Set calibrator to 990 kHz and adjust amplitude output adjustment controls until
TI indication equals TI indication recorded in (5) above. Calibrator Error display indication
will be within 0.7 percent.
(24) Set calibrator to 1.5 MHz and adjust amplitude output adjustment controls until
TI indication equals TI indication recorded in (5) above. Calibrator Error display indication
will be within 3 percent.