TB 9-6625-2324-24
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity. The item selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to one ration
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc voltage:
Fluke, Model 5720A (5720A) (p/o
Range: 1 mV to 2.5 V
MIS-35947); w amplifier, Fluke
Accuracy: 1 mV, 0.375%
5725A/AR (5725A/AR)
1 V, 4%
0.25 and 2.5 V, 1.12%
Ac voltage:
Range:
1 mV to 600 V
Accuracy (%)
Voltage
10 Hz
1.9 mV to 190 V
1.25
40 Hz
1.9 mV
0.75
19 mV
0.5
190 mV to 600 V
0.25
500 Hz
1.9 mV
0.5
100 mV to 100 V
0.125
1 kHz
1 mV
0.5
10 mV
0.25
100 mV to 3 V
0.125
5 kHz
1.9 mV
0.5
19 mV
0.25
190 mV to 190 V
0.125
600 V
0.125
40 kHz
600 V
0.125
50 kHz
100 mV to 100 V
0.125
100 kHz
1.9 mV
1
19 mV
0.25
190 mV to 190 V
0.125
190 kHz
1 mV
1
1.5 MHz
1 mV
1
10 MHz
1 mV
1.25
190 kHz
10 mV
0.25
990 kHz
10 mV
0.5
1.5 MHz
10 mV
1
10 MHz
10 mV
1.25
190 kHz
100 mV to 3 V
0.125
990 kHz
100 mV to 3 V
0.175
1.5 MHz
100 mV to 3 V
0.75
10 MHz
100 mV to 3 V
1.25
Range: -15.2 to +15.1 V dc
Fluke, Model 8840A/AF05
Accuracy:
0.17%
(AN/GSM-64D)