TB 9-6625-2326-24
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
e. Unless otherwise specified, on screen controls will be accessed by using the mouse or
the trackball controls to move the arrow over the icon, virtual instrument controls, etc. and
clicking the left button.
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Insert TI into controller PCMCIA slot.
b. Navigate to and execute DM224 Calibration Software program on controller.
NOTE
During Self Test, test lead connector must be disconnected
from TI.
c. Click on Self Test button in display window. Upon successful completion of Self
Test, reconnect test lead connector to TI and proceed to paragraph 8.
a. Performance Check
(1) Connect TI red (+), black (-), and green (ground) connections to calibrator
OUTPUT terminals.
(2) Click on Performance Check button and follow on-screen instructions.