Table 25. Distortion Accuracy
Pulse Modulation

Calibration Procedure For Signal Generator Model 801.0001.43 Page Navigation
  19    20    21    22    23  24  25    26    27  
img
. . . . . . .
TB 6-6625-2329-24
(15) Set measuring receiver to measure FM and PEAK + with all filters off. After
first FM measurement, set measuring receiver to RATIO and LOG mode.
(16) Set up audio analyzer source frequency for each row in table 27. Measuring
receiver will indicate within limits specified in table 27.
Table 27. Ext Modulation Frequency Response
Audio analyzer
Measuring Receiver
source
indications
frequency
(dB)
(Hz)
Min
Max
100
-0.5
+0.5
1000
-0.5
+0.5
2000
-0.5
+0.5
5000
-0.5
+0.5
10000
-0.5
+0.5
20000
-0.5
+0.5
50000
-0.5
+0.5
(17) Press PARAMETERLEVEL pushbutton and press ON/OFFOFF key.
(18) Reduce output on audio analyzer.
b. Adjustments. No adjustments can be made.
16. Pulse Modulation
a. Performance Check
(1) Connect equipment as shown in figure 7.
Figure 7. Pulse modulation.
(2) Press pulse generator pushbuttons as listed in (a) through (h) below:
(a) CHANNEL A.
25