Phase Modulation
Table 22. Distortion Accuracy

Calibration Procedure For Signal Generator Model 801.0001.43 Page Navigation
  15    16    17    18    19  20  21    22    23    24    25  
img
. . . .
TB 6-6625-2329-24
Table 20. PM Accuracy
Measuring receiver modulation indications
Test instrument
DATA and
(rad)
ENTER/UNITS
(rad)
Min
Max
20
18.9
21.1
10
9.4
10.6
3
2.75
3.25
1
0.85
1.15
.3
0.185
0.415
M pushbutton and press ON/OFF-OFF key.
(6) Press PARAMETER
(7) Press pushbuttons and adjust controls as listed in (a) through (c) below for each
row in table 21 below:
(a) Press TI PARAMETER RF pushbutton, enter DATA numerical value, and
press ENTER/UNITS MHz key.
(b) Press TI PARAMETER  M pushbutton, enter DATA numerical value,
and press ENTER/UNITS rad key.
(c) Using measuring receiver and  M measurement techniques, modulation
will be between specified limits.
Table 21. PM Accuracy
Measuring receiver
Test instrument
modulation indications
(rad)
DATA and
DATA and
ENTER/UNITS
ENTER/UNITS
Min
Max
(MHz)
(rad)
46
5
4.65
5.35
46
2.5
2.275
2.725
93.5
10
9.4
10.6
93.5
5
4.65
5.35
Table 21. PM Accuracy - Continued
Measuring receiver
Test instrument
modulation indications
(rad)
DATA and
DATA and
ENTER/UNITS
ENTER/UNITS
Min
Max
(MHz)
(rad)
187.5
20
18.9
21.1
187.5
10
9.4
10.6
375
40
37.9
42.1
375
20
18.9
21.1
750
80
75.9
84.1
750
40
37.9
42.1
M pushbutton and press ON/OFFOFF key.
(8) Press PARAMETER
(9) Adjust controls as listed in (a) through (b) below for each row in table 22 below: