TB 9-6625-2329-35
Table 9. RF
Attenuator Accuracy (Non-interrupted)
Measuring receiver indications
Test instrument
VARIATION
(dBm)
STEP
(dBm)
Min
Max
-5
-4.8
-5.2
-10
-9.5
-10.5
-15
-14.5
-15.5
-20
-19.5
-20.5
(13) Press SHIFT-INSTR PRESET pushbuttons, PARAMETER-LEVEL
pushbutton, and press ON/OFF-OFF key.
(14) Press measuring receiver SHIFT-INSTR PRESET pushbutton.
(15) Disconnect measuring SENSOR MODULE from TI RF 50Ω.
b. Adjustments. No adjustments can be made.
a. Performance Check
(2) Press SHIFT-INSTR PRESET pushbuttons.
(3) Adjust controls as listed in (a) through (c) below for each row in table 10 below.
(a) Press TI PARAMETER-RF pushbutton, enter DATA numerical value, and
press ENTER/UNITS MHz key.
(b) Press TI PARAMETER-LEVEL pushbutton, enter DATA numerical value
and press ENTER/UNITS dBm key.
(c) Set spectrum analyzer to TI frequency, set power reference then tune to