TB 9-6625-2329-35
(4) Press TI PARAMETERRF pushbutton, enter DATA numerical value, and
press ENTER/UNITS frequency key for each row in table 6. Measure and record the RF
power using the measuring receiver.
(5) Calculate the flatness using the formula below. If calculated flatness is not less
than the maximum limits listed in table 6, perform b below.
FORMULA
Flatness = (highest reading lowest reading)/2
Measuring receiver
Test instrument
DATA and
indications
ENTER/UNITS
MHz
Reading
Maximum limit
(dB)
<(dB)
.1
1
.2
1
.3
1
.4
1
.5
1
.6
1
.7
1
.8
1
.9
1
1
1
2
1
3
1
4
1
5
1
6
1
7
1
8
1
9
1
10
1
20
1
30
1
40
1
50
1
60
1
70
1
80
1
90
1
100
1
200
1
10 CHANGE 1