TB 9-6625-2329-35
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check.
NOTE
Before connecting TI, the protective earth terminal of the
instrument must be connected to the protective conductor of
the line power cord. The line plug shall only be inserted in a
socket outlet provided with a protective earth contact. The
protective action must not be negated by the use of an
extension cord (power cable) without a protective conductor
(grounding). Grounding one conductor of a two-conductor
outlet is not sufficient protection.
NOTE
When indications specified in this procedure are not within
tolerance, perform the power supply check prior to making
adjustments.
CAUTION
Before connecting TI to power source, make sure TI is set to the
power source line voltage as shown on rear of TI.
a.. Connect TI to a 115 V ac source.
b. Set POWER switch to ON and allow at least 1 hour for TI to warm-up and stabilize.
c.
Connect equipment as
shown in
figure 1 below: