TB 9-6625-2329-35
d. In
figure 2 adjust REF FREQ 10 MHz for minimum drift
indication on the
frequency difference meter.
e.
Check frequency drift
after 24
hours. The drift will be
less than 1 part in 10
9.
f.
Disconnect frequency difference meter from TI and
time/frequency workstation.
a. Performance Check
(1) Connect equipment as shown in figure 3 below: