TB 9-6625-2330-35
(j)
Using rotary knob, adjust delta marker to bottom of square wave.
(6) Using spectrum analyzer, measure top to bottom of square wave in dB. Pulse
envelope on/off ratio will indicate within limits specified in table 15.
Spectrum analyzer
>dB
60
(7) Set TI FUNCTION-RF ON to off (red light extinguished).
(9) Adjust pulse generator output for a period of 20 s and a pulse width of 15 s.
(10) Set TI FUNCTION-RF ON to on (red light lit).
(11) Set oscilloscope input impedance to 50 Ω.
(12) Using oscilloscope measurement techniques, verify risetime of displayed
envelope is within limits listed in table 16.
Oscilloscope
< ns
25
(13) Using oscilloscope measurement techniques, verify falltime of displayed
envelope is within limits listed in table 17.
Oscilloscope
< ns
25
(14) Set TI FUNCTION-RF ON to off (red light extinguished).