(2) Press TI pushbuttons as listed in (a) through (e) below:
(a)
FUNCTION-CARR FREQ.
(b)
DATA ENTRY- 500 MHz.
(c)
FUNCTION-CARR LEVEL.
(d)
DATA ENTRY- 0 dBm.
(e)
FUNCTION-RF ON to on (red light lit).
(3) Using measuring receiver and RF power measurement techniques, measured
power will indicate within limits specified in table 10 for 0 dBM.
(4) Set measurement receiver to reference mode.
(5) Set TI to next power level indicated in table 10.
(6) Using standard tuned level measurement techniques, measuring receiver will
indicate within limits specified for TI output level in table 10.
(7) Repeat (5) and (6) above for remaining output levels listed in table 10.
Measuring receiver
Test instrument
indications
DATA ENTRY
(dB)
level
Min
Max
0 dBm
-1.5
1.5
-10 dBm
-11.5
-8.5
-20 dBm
-21.5
-18.5
-30 dBm
-31.5
-28.5
-40 dBm
-41.5
-38.5
-50 dBm
-51.5
-48.5
-60
dBm
-61.5
-58.5
-70
dBm
-71.5
-68.5
-80
dBm
-81.5
-78.5
-90
dBm
-91.5
-88.5
-100
dBm
-101.5
-98.5
-110
dBm
-111.5
-108.5
(8) Press TI pushbuttons as listed in (a) through (e) below:
(a)
FUNCTION-CARR FREQ.
(b)
DATA ENTRY-1300 MHz.
(c)
FUNCTION-CARR LEVEL.
(d)
DATA ENTRY- 0 dBm.
(e)
FUNCTION-RF ON to on (red light lit).
(9) Using measuring receiver and RF power measurement techniques, measured
power will indicate within limits specified in table 11 for 0 dBM.
(10) Set measurement receiver to reference mode.
(11) Set TI to next power level indicated in table 11.
(12) Using standard tuned level measurement techniques, measuring receiver will
indicate within limits specified for TI output level in table 11.