TB 9-6625-2330-35
(c) FUNCTION-CARR LEVEL.
(d) DATA ENTRY- 13 dBm.
(e) FUNCTION-RF ON to on (red light lit).
(12) Using RF power measurement techniques, measuring receiver will indicate
within limits specified in table 7 for TI RF power level setting.
(13) Press TI FUNCTION-CARR LEVEL and DATA ENTRY keys to set TI output
level to next value listed in table 7. Measuring receiver will indicate within limits specified
(14) Repeat (13) above for remaining levels listed in table 7.
Measuring receiver
Test instrument
indications
DATA ENTRY
(dB)
level
Min
Max
13
dBm
10.5
15.5
10
dBm
7.5
12.5
5
dBm
2.5
7.5
0
dBm
-2.5
2.5
(15) Press TI FUNCTION-CARR LEVEL and DATA ENTRY keys to set TI output
(16) Using standard tuned level measurement techniques, measuring receiver will
indicate within limits specified in table 8 for TI RF level setting.
(17) Press TI FUNCTION-CARR LEVEL and DATA ENTRY keys to set TI output
level to next value listed in table 8. Measuring receiver will indicate within limits listed in
(18) Repeat (17) above for remaining levels listed in table 8.
Measuring receiver
Test instrument
indications
DATA ENTRY
(dB)
level
Min
Max
0 dBm
-2.5
2.5
-10 dBm
-12.5
-7.5
-20 dBm
-22.5
-17.5
-30 dBm
-32.5
-27.5
-40 dBm
-42.5
-37.5