Figure 5. Trigger level adjust.

Calibration Procedure For Pulse Generator (Model 9210 Mod 200) Page Navigation
  17    18    19    20    21  22
TB 9-6625-2332-40
(6) Verify for each positive pulse displayed on Vertical 1 of the oscilloscope that
there is one positive pulse displayed on Vertical 2.
(7) Press TI pushbuttons as listed in (a) and (b) below:
(a) Delay and enter 200 from data keyboard and press N/GHz.
(b) 2 pulse On.
(8) Verify for each positive pulse displayed on Vertical 1 of the oscilloscope that
there are two positive pulses displayed on Vertical 2.
(9) Disconnect all equipment from the TI.
b. Adjustments. No adjustments can be made.
16. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
23/(24 Blank)