Delay
Figure 4. Trigger level and polarity.

Calibration Procedure For Pulse Generator (Model 9210 Mod 200) Page Navigation
  12    13    14    15    16  17  18    19    20    21    22  
img
. . .
TB 9-6625-2332-40
Table 16. 9211 Delay
Digitizing oscilloscope
Digitizing oscilloscope
T indications (sec)
Test instrument settings
settings
Per
Wid
Delay
Time range
Time delay
Min
Max
20  n
17
n
10  n
20  nsec
30  nsec
8.95  nsec
11.05  nsec
200  n
175
n
100  n
200  nsec
190  nsec
98.5
nsec
101.5
nsec
sec
1.8  sec
0.994  sec
1.006  sec
2
1.75
1
2
sec
sec
9.949  sec
10.051  sec
20
17.5
10
20
18
sec
sec
sec
sec
200
175
100
200
180
99.5
100.5
2m
1.75 m
1m
2  msec
1.8 msec
0.995 msec
1.005 msec
20  m
17.5  m
10  m
20  msec
18  msec
9.95  msec
10.05  msec
200  m
175
m
100  m
200  msec
180  msec
99.5
msec  100.5
msec
(15) Press DISABLE pushbutton (model 9211 output module) to off (red light lit).
(16) Disconnect cable from TI 9211 module OUTPUT.
(17) Connect cable to TI 9215 module OUTPUT.
(18) Press TI pushbuttons as listed in (a) through (h) below:
(a) CHANNEL B.
(b) Vhigh and enter 5 from data keyboard and press ENTER/Hz.
(c) Vlow and enter 0 from the data keyboard and press ENTER/Hz.
(d) Lead and enter 5 from data keyboard and press n/GHz.
(e) Trail and enter 5 from data keyboard and press n/GHz.
(f) Per and enter 40 from data keyboard and press n/GHz.
(g) Wid and enter 10 from data keyboard and press n/GHz.
(h) Press DISABLE pushbutton (model 9215 output module) to on (red
light extinguished).
(19) Setup digitizing oscilloscope to display the pulse train.
(20) Setup digitizing oscilloscope to perform a T measurement by placing V marker
1, V marker 2, Start Marker and Stop Marker at the same point of the leading edge of the
left most pulse that is displayed on the crt.
(21) Press TI pushbuttons as listed in (a) and (b) below:
(a) CHANNEL B.
(b) Delay and enter 10 from data keyboard and press n/GHz.
(22) Set digitizing oscilloscope Stop marker on the leading edge of the displayed pulse
where the V markers intersect the pulse.
(23) Verify that the digitizing oscilloscope T indication is within the limits listed in
table 17 for the Delay setting.
(24) Set TI to the next Per and Wid settings listed in table 17.
(25) On the digitizing oscilloscope turn off the V markers and the Start and Stop
markers.
(26) Repeat (22) and (23) above.
18