TB 9-6625-2332-40
b. Adjustments. No adjustments can be made.
a. Performance Check
(2) Reset digitizing oscilloscope and cal the gain.
(3) Press TI pushbuttons as listed in (a) through (h) below:
(a) CHANNEL A.
(b) Vhigh and enter 2.5 from data keyboard and press ENTER/Hz.
(c) Vlow and enter 0 from the data keyboard and press ENTER/Hz.
(d) Lead and enter .9 from data keyboard and press n/GHz.
(e) Trail and enter .9 from data keyboard and press n/GHz.
(f) Per and enter 20 from data keyboard and press n/GHz.
(g) Wid and enter 17 from data keyboard and press n/GHz.
(h) Press DISABLE pushbutton (model 9211 output module) to on (red light
extinguished).
(4) Setup digitizing oscilloscope to display the pulse train.
(5) Setup digitizing oscilloscope to perform a t measurement by placing V marker
1, V marker 2, Start Marker and Stop Marker at the same point of the leading edge of the
left most pulse that is displayed on the crt.
(6) Press TI pushbuttons as listed in (a) and (b) below:
(a) CHANNEL A.
(b) Delay and enter 10 from data keyboard and press n/GHz.
(7) Set digitizing oscilloscope Stop marker on the leading edge of the displayed pulse
where the V markers intersect the pulse.
(8) Verify that the digitizing oscilloscope T indication is within the limits listed in
(9) Set TI to the next Per and Wid settings listed in table 16.
(10) On the digitizing oscilloscope turn off the V markers and the Start and
Stop markers.
(11) Repeat (4) and (5) above.
(12) Press TI pushbuttons as listed in (a) and (b) below:
(a) Channel A.
(b) Delay and enter the next delay value in table 16 from the data keyboard.
(13) Repeat (7) and (8) above.
(14) Repeat (9) through (13) above for the remaining Per, Wid and Delay settings