TB 9-6625-2332-50
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
was used as the prime data source in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time
required for this
calibration is
approximately 6
hours, using the dc and low frequency and microwave technique.
2. Forms,
Records, and
Reports. Forms, records, and
reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
9210 Main Frame
Period accuracy
Range: 1 nsec to 500 msec
Accuracy: 0.5% of programmed value + 0.2 nsec
Pulse width accuracy
Range: 1 nsec to 500 msec
Accuracy: 0.5% of programmed value + 0 nsec
Pulse delay accuracy
Range: 0 nsec to 500 msec
Accuracy: 0.5% of programmed value + 1 nsec
≤0.035% of programmed value + 5 psec
RMS jitter
Accuracy: 100 mV
External trigger
Trigger Slope: Positive or negative
Double Pulse: When double pulse is set to on, two pulses are
produced for each trigger pulse.
9211 Module
Range: 5 V
Amplitude
Accuracy: 1% of programmed value +1% of amplitude +40 mV
into 50Ω
Variable risetime/falltime
10% to 90% point
Accuracy: 10% of programmed value + 300 psec
9215 Module
Range: 15 V
Amplitude
Accuracy: 1% of programmed value + 1% of amplitude + 5 mV
into 50Ω
Variable risetime/falltime
Range:
10% to 90% point
Accuracy: 20% of programmed value + 300 psec